Ethylene oxide sterilization: packaging, indicators and aeration

Which reels, pouches and wraps suit ethylene oxide sterilization, which indicators to use inside and outside packs, and why aeration time matters.

In short: ethylene oxide (EO) sterilizes heat- and moisture-sensitive devices at low temperature. It needs packaging that lets the gas in and out — paper-film or Tyvek®-film reels and pouches, or crepe wrap — EO-specific indicator tape and in-pack indicators, and a full aeration time before use to remove residual gas.

How EO sterilization works

Ethylene oxide is an alkylating gas that inactivates microorganisms at around 30–54 °C with controlled humidity. It penetrates packaging and long lumens well, which makes it suitable for complex, heat-sensitive devices. Its drawbacks are a long total cycle — conditioning, exposure and aeration can take many hours — and toxicity, so residues must fall below the limits in ISO 10993-7 before devices are used.

Packaging for EO

PackagingSuitable for EONotes
Paper-film reels and pouchesYesThe most common choice; check the printed EO indicator
Tyvek®-film reelsYesStrong and lint-free; also suits H₂O₂ plasma
Crepe wrapping paperYesFor trays and bulky sets
Rigid containersOnly if validated for EOCheck the manufacturer’s instructions
Foil, PVC, nylon filmNoBlock gas penetration or absorb EO

Indicators for EO

  • Outside: EO indicator tape or the printed EO indicator on reels — Class 1, shows the pack was processed.
  • Inside: an EO-specific in-pack chemical indicator in each pack.
  • Biological: EO cycles are challenged with Bacillus atrophaeus spores — see biological indicators.

Steam indicators do not respond correctly to EO, and EO indicators do not respond to steam; keep them separate in the store.

Preparing an EO load

  1. Clean and dry devices thoroughly; water reacts with EO and reduces effectiveness.
  2. Remove caps and stylets so gas reaches every surface, unless the device instructions say otherwise.
  3. Pack loosely — do not overfill pouches or stack packs tightly.
  4. Label on the film side or on the label area.
  5. After the cycle, aerate for the full time the sterilizer and device manufacturers specify before release.

Staff safety

EO is toxic and flammable. Sterilizers must be installed with exhaust ventilation and leak monitoring, and staff should handle loads only after aeration. Follow the sterilizer manufacturer’s safety instructions and local occupational-health rules.

For low-temperature alternatives, compare hydrogen-peroxide plasma.