GClasses & pathways guide
DRAP medical device classes, enlistment and registration: a working guide
How the Medical Devices Rules 2017 decide what a device needs before it can be imported, made or sold in Pakistan — the classes, the two routes, who may apply, what goes in the file, what it costs, how long it lasts and what the holder owes DRAP afterwards.
In short. Every medical device sold in Pakistan needs two things from DRAP: a licensed Pakistani establishment behind it, and its own certificate. Class A devices are enlisted; Class B, C and D devices are registered. Both certificates last five years, are filed online on DRAP's MDMC portal, and must be current before PSW will accept an import.
Who regulates medical devices in Pakistan
The Drug Regulatory Authority of Pakistan (DRAP) regulates devices through its Medical Devices & Medicated Cosmetics (MDMC) Division under the DRAP Act 2012 and the Medical Devices Rules 2017, notified as S.R.O. 32(I)/2018. Licences, enlistments, registrations and import permits are decided by the Medical Device Board (MDB). In-vitro diagnostic devices, software, implants and accessories all fall within the rules' definition of a medical device.
A licence for the firm, a certificate for the device
DRAP authorises the business and the product separately. An establishment licence lets a firm import (ELI, applied for on Form-2) or manufacture (ELM, Form-1), and covers its storage premises — which an inspection panel visits against DRAP's good storage practice checklist. Good distribution practice for medical devices is a licence condition. The licence authorises no product: each device still needs its own enlistment or registration.
- Gate 1Establishment licenceOnce per firm · 5 years
- Gate 2Enlistment or registrationPer device or group · 5 years
- Gate 3PSW import applicationPer consignment
- Gate 4Goods declaration & releasePer consignment
The four classes
Devices are classed by risk from A (lowest) to D (highest) using the classification rules in Schedule A, which weigh how invasive the device is, how long it stays in contact with the body, whether it is active, and whether it carries a medicine or tissue. The class follows the manufacturer's stated intended use — the same product with a different claim can move class.
| Class | Risk | Examples | Route | Form |
|---|---|---|---|---|
| A | Low | Tongue depressors, reusable manual instruments | Enlistment | Form-6A import · Form-6 local |
| B | Low–moderate | Hypodermic needles, suction equipment | Registration | Form-7A import · Form-7 local |
| C | Moderate–high | Lung ventilators, bone fixation plates | Registration | Form-7A import · Form-7 local |
| D | High | Heart valves, implantable defibrillators | Registration | Form-7A import · Form-7 local |
Several devices can travel on one application when Schedule B allows them to be grouped as a single device, family, set, system, kit or cluster. Grouping well is the largest single saving in a registration programme.
Enlistment vs registration
| Enlistment | Registration | |
|---|---|---|
| Classes | A | B, C and D |
| Imported devices | Form-6A | Form-7A |
| Locally made devices | Form-6 | Form-7 |
| Depth of review | Administrative and conformity evidence | Full technical evaluation, with queries |
| Validity | 5 years | 5 years |
| Decided by | Medical Device Board, on DRAP's MDMC portal | |
Who may apply — and the authorized representative
Rule 14 allows two applicants: a licensed Pakistani manufacturer for its own products, or an importer holding an establishment licence, acting as the foreign manufacturer's authorized representative. A manufacturer abroad cannot file on its own. The importer's name and address must then appear on the label of every foreign-made device (rule 42), and the representative carries the DRAP correspondence and vigilance duties for the certificate's life. Changing representative is a change of particulars that needs the Board's prior approval — how ProDr acts as representative.
What goes in the dossier
Application form, letter of authorisation from the manufacturer, the applicant's establishment licence, fee challan.
ISO 13485 certificate, CE certificate or another reference approval, declaration of conformity, Free Sale Certificate attested by the Pakistan mission in the country of manufacture.
Device description and intended use, classification rationale, specifications, essential-principles checklist, risk management, design verification and validation; clinical evidence and stability data as the class requires.
Label artwork and IFU, proposed MRP, shelf life and storage conditions.
Approvals from the reference authorities named in rule 67 — the USA, Japan, Australia, Canada, the UK, Switzerland, Norway and listed EU states — carry the most weight, and can waive a foreign site inspection, as can WHO prequalification or CE marking by a NANDO-listed notified body.
Government fees
Set by S.R.O. 1399(I)/2025, effective 1 August 2025. Fees are per application, non-refundable, and paid by challan before filing. Service fees of any consultant or representative are separate.
| Application | New | Renewal |
|---|---|---|
| Class A enlistment — import or local | 6,300 | Included |
| Class B registration — import | 32,400 | 15,700 |
| Class C or D registration — import | 64,800 | 32,400 |
| Class B, C or D registration — local manufacture | 26,100 | 12,500 |
| Establishment licence to import (ELI) | 26,100 | 12,500 |
| Establishment licence to manufacture (ELM) | 128,500 | 64,800 |
| Change of particulars | 50% of the relevant fee | |
| Appeal to the Appellate Board | 64,800 | — |
Validity, renewal and lapse
Enlistment and registration certificates run five years from issue unless suspended or cancelled; establishment licences follow a parallel clock. Renewal is filed 60 days before expiry. A late renewal remains possible for up to one year after expiry, with an extra fee equal to the renewal fee for each month of delay. After that year the authorisation ceases and only a fresh application remains.
Changes after approval
Any change in particulars — manufacturer, site, label, representative — and any change that affects safety or performance is a prior-approval variation on the original form, at half the relevant fee. The changed device may not be made or placed on the market until the Board approves it.
Labelling
Labels must be in English, and also in Urdu for devices intended for home use (rule 41). They carry the device name and identifiers, the enlistment or registration number, lot or serial number, dates of manufacture and expiry, the MRP, and the manufacturer's details with licence number — plus the importer's name and address for foreign-made devices (rule 42). The IFU sets out contraindications, warnings and precautions (rule 44). Device advertisements need DRAP approval, valid for two years and limited to the authorised purpose.
Vigilance clocks
| Event | Deadline | Rule |
|---|---|---|
| Batch failure, unexpected side effect, injury, toxicity or performance failure | 14 days | 19(9)(b) |
| Safety withdrawal, ban or restriction in a reference country | 30 days | 19(10) |
| Adverse reaction not in the application, reported in a reference country | 30 days | 19(12) |
| Reply to a show-cause notice | 15 days | 10(2), 20(2) |
| Appeal against a Board decision | 60 days | 60 |
Importing through PSW
No device clears customs without DRAP's clearance (rule 26). DRAP's letter of 16 October 2026 directs medical device import applications to the Pakistan Single Window, keeping its older Online Import & Export System only for importers who cannot yet file through PSW. PSW processes an import only when DRAP's system holds the importer's current licence and the device's current enlistment or registration, so importers were asked to complete their data update in DRAP's licensing and registration system. Radiation-emitting equipment also needs PNRA authorisation.
Special routes
Donated devices may be imported, except Class C and D devices and Schedule D products, and may never be sold.
Small quantities for clinical investigation, examination, test or analysis, permitted on Form-10.
The Board may authorise emergency use during a declared public-health emergency, disaster or pandemic.
Since 1 September 2026, refurbished devices need their own enlistment or registration under S.R.O. 1450(I)/2026. Single-use, implantable and disposable devices may never be refurbished.
Key dates
- Medical Devices Rules 2017 notifiedS.R.O. 32(I)/2018
- Transition ends for Classes C and DS.R.O. 224(I)/2023
- Transition ends for Classes A and BCertificates required for every class
- MDMC portal goes livePaper applications no longer accepted
- Revised feesS.R.O. 1399(I)/2025
- Refurbishment RegulationsS.R.O. 1450(I)/2026
- Device imports through PSWOIES kept only as a fallback
Frequently asked
Which DRAP division regulates medical devices?
The Medical Devices & Medicated Cosmetics (MDMC) Division. Applications are decided by the Medical Device Board (MDB).
Which countries count as reference countries for DRAP?
Rule 67 names the USA, Japan, Australia, Canada, the UK, Switzerland, Norway, Austria, Belgium, Denmark, France, Germany, Ireland, Italy, the Netherlands, Spain and Sweden. WHO-prequalified devices and CE-marked devices certified by a NANDO-listed notified body are also relied on, subject to evidence.
Do device labels need to be in Urdu?
English is required on every label. Urdu is additionally required for devices intended for home use (rule 41).
How quickly must adverse events be reported to DRAP?
Batch failures, unexpected side effects, injury, toxicity and performance failures go to the Medical Device Board within 14 days. Safety withdrawals, bans or restrictions in reference countries must be reported within 30 days.
Sources
- Medical Devices Rules 2017, as amended to April 2022 — DRAP
- Revised regulatory fee for therapeutic goods, S.R.O. 1399(I)/2025 — DRAP
- S.R.O. 224(I)/2023, amendment to the Medical Devices Rules — DRAP
- Guidelines on Good Distribution Practice for Medical Devices — DRAP
- DRAP e-services and MDMC portal
This guide summarises the rules for planning. It is not legal advice; confirm the current form, fee and procedure with DRAP or with our desk before filing.