Reading the ECG paper grid: speed, gain and the calibration pulse
What each small and large square means at 25 mm/s and 10 mm/mV, how speed and gain changes alter the trace, and quick heart-rate and interval checks.
In short: standard ECG paper runs at 25 mm per second with a gain of 10 mm per millivolt. At those settings, one small square (1 mm) is 0.04 seconds wide and 0.1 mV high, and one large square (5 mm) is 0.2 seconds and 0.5 mV. Always check the speed, gain and calibration pulse printed on the trace before measuring anything.
The grid at standard settings
| Size | Time (across) | Voltage (up) | |
|---|---|---|---|
| Small square | 1 mm | 0.04 s | 0.1 mV |
| Large square | 5 mm (5 small) | 0.2 s | 0.5 mV |
| Five large squares | 25 mm | 1 s | — |
| Two large squares (up) | 10 mm | — | 1 mV |
The calibration pulse
Most electrocardiographs print a square calibration pulse at the start of each line or page. At the standard gain it is 10 mm tall — two large squares — showing that 1 mV equals 10 mm. If the pulse is a different height, the gain has been changed.
- Half gain (5 mm/mV): used when very tall QRS complexes overlap neighbouring leads. Every voltage on the trace must be doubled when measured.
- Double gain (20 mm/mV): used for small complexes. Voltages must be halved.
When the speed changes
At 50 mm/s, often used for children and to look at fast rhythms, everything is twice as wide: one small square is 0.02 seconds. A normal QRS then looks broad to an eye used to 25 mm/s. Check the speed printed on the report before calling an interval long or short. CTG paper uses different speeds again — see CTG paper speed.
Quick measurements at 25 mm/s
- Heart rate, regular rhythm: 300 ÷ the number of large squares between two R waves, or 1,500 ÷ the number of small squares.
- Heart rate, irregular rhythm: count the QRS complexes on a 10-second strip (50 large squares) and multiply by 6.
- PR interval: normally 0.12–0.20 s, or 3 to 5 small squares.
- QRS duration: normally under 0.12 s, or fewer than 3 small squares.
Why the paper matters
Most 12-channel machines print on paper with a pre-printed grid. If the grid is not printed to the right pitch, or the cue marks that tell the machine where each page starts are in the wrong place, the trace and grid drift apart and measurements become unreliable. Use paper made for your machine model, load it the right way up (thermal side towards the print head), and store it away from heat and light so the grid and trace stay readable — see storing thermal paper.
EchoGrand™ recording paper is matched to the grid, size and cue marks of each machine it is listed for. Find yours in the ECG paper size guide or browse all recording paper.