+92 318 5412255 +92 51 2769 848 info@prodr.com.pk Delivering nationwide
Partner desk Enquiry basket0 Contact Us

CTG paper speed: 1, 2 or 3 cm per minute — and why it matters

How CTG paper speed (1, 2 or 3 cm/min) changes how fetal heart rate traces look, how to match CTG paper to the monitor and scale, and labour-ward good practice.

In short: CTG traces are printed at 1, 2 or 3 cm per minute, and the paper grid must match the monitor’s speed and heart-rate scale. Reading a trace printed at a different speed from the one you trained on changes how variability and decelerations look. Confirm the speed printed on every trace, and order paper that matches the monitor model.

Why paper speed matters

At 3 cm/min, one minute of fetal heart rate takes three centimetres of paper; at 1 cm/min it takes one. The same trace therefore looks compressed at 1 cm/min and stretched at 3 cm/min. Baseline variability can look reduced at the faster speed and exaggerated at the slower one, and decelerations change shape. Clinicians are trained to read at one speed, so a mismatch is a recognised source of misinterpretation.

Common settings

SpeedWhere it is commonPaper per hour
1 cm/minUK and much of Europe60 cm
2 cm/minSome European and Asian units120 cm
3 cm/minUSA and many other countries180 cm

Hospitals in Pakistan use monitors set to different speeds depending on the brand and the unit’s training, so check rather than assume.

Matching paper to the monitor

  • Model and part number. Order by monitor model or the OEM paper number — for example M1911A paper for Philips Avalon fetal monitors.
  • Heart-rate scale. Pre-printed grids use a set bpm range and scale (for example 30 or 50–210 bpm). The paper must match the scale the monitor is configured for.
  • Grid printed by the monitor. Some monitors, such as the Sonicaid 8000 series, print their own grid on plain thermal paper; the paper then only has to match width, fold and sensing marks.
  • Z-fold format. CTG paper is supplied in Z-fold packs so long traces can be reviewed page by page and filed.

Good practice on the labour ward

  1. Check the paper speed and the date and time printed on the trace at the start of monitoring.
  2. Label each trace with the patient’s details and sign any event marks.
  3. Do not change the speed mid-trace unless your protocol requires it; if it changes, annotate it.
  4. Store traces away from heat, light and plastic sleeves — see why thermal traces fade.

For the basics of the two traces, see cardiotocography; for every paper format by machine, see the paper size guide.