Biological vs chemical indicators: what each proves in sterilization

How physical, chemical and biological monitoring differ, which spores test which process, how often to run BIs, and what to do when one fails.

In short: a chemical indicator shows that a pack saw the conditions of a cycle; a biological indicator shows that the cycle actually killed a large population of highly resistant spores. Chemical indicators give an answer at once, for every pack. Biological indicators take time to read and are used on a schedule — routinely at least weekly, and in every load that contains implants. A CSSD needs both, together with the sterilizer’s own physical records.

Three layers of monitoring

PhysicalChemicalBiological
What it isThe sterilizer’s printout or data: time, temperature, pressureInk or chemistry that changes colour under process conditionsA carrier of bacterial spores, incubated after the cycle
What it provesThe chamber reached the set valuesThe conditions reached the indicator’s locationThe process killed resistant spores at that location
ResultAt the end of the cycleAt the end of the cycleAfter incubation; hours for rapid-readout types, longer for conventional ones
StandardSterilizer standard and validationISO 11140-1ISO 11138 series

Chemical indicators

Chemical indicators are grouped into classes. Class 1 tape on the outside shows a pack has been through a process. Class 2 is the daily Bowie-Dick test. Classes 4, 5 and 6 go inside packs and respond to more than one variable; Class 5 integrating indicators are designed to correlate with a biological indicator’s response. See indicator classes explained for each class.

A chemical indicator that has changed colour does not prove sterility. It is strong evidence the pack was exposed to the right conditions, and a failed indicator is a clear reason not to use the pack.

Biological indicators

A biological indicator carries a known population of spores chosen because they are hard to kill by that process:

  • Steam and hydrogen-peroxide plasma: Geobacillus stearothermophilus.
  • Ethylene oxide and dry heat: Bacillus atrophaeus.

After the cycle, the indicator is incubated. If the spores grow, the media changes colour (or a reader detects fluorescence in rapid-readout types), and the cycle has failed. An unprocessed indicator from the same lot is incubated alongside as a control, to prove the spores were alive.

When to use each

SituationWhat to use
Every packClass 1 process indicator outside; Class 4, 5 or 6 indicator inside
Every day, pre-vacuum steam sterilizerBowie-Dick test before the first load
Routine efficacy monitoringBiological indicator in a process challenge device — at least weekly, preferably every day the sterilizer is used (AAMI ST79 recommendation)
Loads containing implantsBiological indicator with a Class 5 integrating indicator in every load; hold the implants until the result is known wherever possible
After installation, relocation or major repairBiological indicator tests, as part of requalification

Your national guidance and the sterilizer manufacturer may set different frequencies; follow your local policy.

If a biological indicator fails

  1. Take the sterilizer out of use and tell the CSSD manager and infection control.
  2. Check the physical records and chemical indicators for that load.
  3. Recall items from the failed load, and from earlier loads back to the last passed biological test, as your policy requires.
  4. Find and fix the cause, then retest — usually with consecutive passed biological tests — before releasing loads again.

Good pack labelling and load records make a recall quick and precise.

SteriCraft® covers the chemical layer: steam indicator strips in Class 4, 5 and 6, plasma indicator strips, indicator tapes and the Bowie-Dick test pack.