Sterile pack labelling and load traceability in the CSSD
What to print on every sterile pack label, what to keep in the load record, how to link packs to patients and how to recall a sterilizer load.
In short: every sterile pack should be traceable to the sterilizer, cycle and date it was processed, and every patient use should be traceable back to the pack. A simple label — contents, sterilizer number, load number, date and the packer’s initials — plus a load record that holds the cycle printout and indicator results is enough to recall a load if anything goes wrong.
What goes on the pack label
- Contents or set name.
- Sterilizer number and cycle (load) number.
- Sterilization date (and expiry or rotation date if your policy uses one).
- Initials or ID of the person who packed it.
- Department the pack belongs to, for loan and ward sets.
Write on the label or the film side of reels and pouches, never on the paper side: ink can penetrate the paper and pens can puncture it. Labels must survive the process, so use labels and inks made for the sterilization method.
What goes in the load record
| Record | Why it matters |
|---|---|
| Sterilizer printout or data log (time, temperature, pressure) | Physical proof that the cycle parameters were met |
| Daily Bowie-Dick or air-removal test result | Shows the sterilizer was fit to run that day |
| Load contents list | Lets you find every pack from the load in a recall |
| Chemical indicator results, including any PCD | Shows conditions inside the load |
| Biological indicator results where used | Required for implants and validation by most policies |
| Operator who released the load | Accountability for release decisions |
Linking packs to patients
In theatre, record the pack label — or stick a peel-off copy — in the patient’s notes or the theatre register. Many hospitals use a two-part label for this. Without it, a recall can only say which packs were affected, not which patients.
Recalling a load
- Identify the sterilizer, load number and date from the failed record.
- Quarantine all unused packs from that load, wherever they are.
- Check theatre records for packs already used and inform infection control.
- Find the cause, correct it and document the investigation before the sterilizer returns to service.
Supply traceability
Traceability starts before the CSSD. Record the batch number and expiry of the packaging and indicators used, so a supplier recall can be matched to your loads. ProDr delivery notes list batch and expiry for every line — see compliance and documentation.