Sterile pack labelling and load traceability in the CSSD

What to print on every sterile pack label, what to keep in the load record, how to link packs to patients and how to recall a sterilizer load.

In short: every sterile pack should be traceable to the sterilizer, cycle and date it was processed, and every patient use should be traceable back to the pack. A simple label — contents, sterilizer number, load number, date and the packer’s initials — plus a load record that holds the cycle printout and indicator results is enough to recall a load if anything goes wrong.

What goes on the pack label

  • Contents or set name.
  • Sterilizer number and cycle (load) number.
  • Sterilization date (and expiry or rotation date if your policy uses one).
  • Initials or ID of the person who packed it.
  • Department the pack belongs to, for loan and ward sets.

Write on the label or the film side of reels and pouches, never on the paper side: ink can penetrate the paper and pens can puncture it. Labels must survive the process, so use labels and inks made for the sterilization method.

What goes in the load record

RecordWhy it matters
Sterilizer printout or data log (time, temperature, pressure)Physical proof that the cycle parameters were met
Daily Bowie-Dick or air-removal test resultShows the sterilizer was fit to run that day
Load contents listLets you find every pack from the load in a recall
Chemical indicator results, including any PCDShows conditions inside the load
Biological indicator results where usedRequired for implants and validation by most policies
Operator who released the loadAccountability for release decisions

Linking packs to patients

In theatre, record the pack label — or stick a peel-off copy — in the patient’s notes or the theatre register. Many hospitals use a two-part label for this. Without it, a recall can only say which packs were affected, not which patients.

Recalling a load

  1. Identify the sterilizer, load number and date from the failed record.
  2. Quarantine all unused packs from that load, wherever they are.
  3. Check theatre records for packs already used and inform infection control.
  4. Find the cause, correct it and document the investigation before the sterilizer returns to service.

Supply traceability

Traceability starts before the CSSD. Record the batch number and expiry of the packaging and indicators used, so a supplier recall can be matched to your loads. ProDr delivery notes list batch and expiry for every line — see compliance and documentation.