Sterilization & CSSD
Emulating indicator (Class 6)
Also called: Class 6 chemical indicator, cycle verification indicator
Emulating indicator (Class 6) — A Class 6 chemical indicator designed to react to all critical variables of one specific cycle, for example 134 °C for 3.5 minutes.
Class 6 indicators are cycle-specific: a 134 °C indicator is made for that cycle and is not a substitute for monitoring a 121 °C cycle. They have tight tolerances, so a pass is strong evidence that the specified conditions were met inside the pack.
Choose the Class 6 indicator that matches the validated cycle your sterilizer runs, and record the result with the load.
Related termsChemical indicatorIntegrating indicator (Class 5)Steam sterilization
Questions about Emulating indicator (Class 6)?The ProDr desk answers technical and documentation questions within one business day.
WhatsApp the desk